Hipot Testers & Electrical Safety: V10X Series
V103
- AC Hipot 20V – 6KV / 50mA
- DC Hipot 50V – 11KV / 30mA
- Leakage Resolution 100 pA
- IR Range 6TΩ Range
- Ground Bond Not included
- External Option 30KVAC Available
The V103 extends DC hipot voltage to 11KV, covering high-voltage power supplies, motor drives, industrial equipment, and EV battery systems where 6KV DC testing is insufficient. Full V10X touchscreen and automation architecture throughout.
At 11KV DC with 30mA sourcing current, the V103 covers demanding insulation testing requirements for high-voltage industrial and automotive applications. The same 100 pA leakage resolution and TΩ insulation resistance range as lower-voltage models, extended to higher voltage classes.
Phase angle measurement separates resistive from capacitive leakage, particularly useful at higher voltages where capacitive coupling becomes a more significant factor in insulation assessment. High-power DC output enables rapid charge/discharge cycles for high-capacitance devices such as DC-link capacitors and solar panels.
All V10X automation features are present: graphical touchscreen workflow, real-time charts, built-in PDF and CSV reports, SCPI interfaces over Ethernet, USB, RS232, and optional GPIB. Direct control of up to 1,600 scanner channels.
Overview
The V103 pushes DC hipot to 11KV in the V10X platform, for industrial power equipment, EV components, and high-voltage cable systems requiring more than standard 6KV DC testing. All V10X automation and measurement capabilities included.
Specifications
- AC Hipot: 20V – 6KV, 50mA (100mA / 200mA options)
- DC Hipot: 50V – 11KV, 30mA
- IR Range: 6TΩ range
- Leakage resolution: 100 pA
- Phase angle measurement: In-phase and quadrature components
- Arc detection: 4–40µS / 1–30mA, configurable
- Breakdown detection: <150µS
- HV scanner control: Up to 1,600 channels
- 30KVAC external option available
Additional Information
- Interfaces: Ethernet/LAN, USB, RS232, Digital I/O, optional GPIB, embedded WiFi
- Reports: Built-in PDF + CSV generation
- Certification: EN 61010-1 / EN 61326-1
- Made in USA
RELATED PRODUCTS
While you are here
964i
964i High Voltage Switching System- Max Voltage Up to 15KV DC / 10KV AC
- Max Current 40A (HC Series cards)
XiTRON XT2640
XiTRON XT2640 Multi-Channel Harmonic Power Analyzer- Power Accuracy 0.03%
- Virtual Power Analyzers Up to 3 (in single chassis)
PA900
PA900 Precision Multi-Channel Harmonic Power Analyzer- Power Accuracy 0.03%
- Bandwidth 1 MHz
Showing 4 of 16 Hipot Testers & Electrical Safety: V10X Series
Request a quote to see the full catalogue >